Chapter 2: Control Solution Testing
Why Run A Control Solution Test
We recommend that you run the EasyMax V Normal control test because it lets you know that your meter and
test strips are working properly to give you reliable results. You should run the control solution test when:
You use the EasyMax V Blood Glucose Meter for the first time.
You open a new bottle of test strips.
You think the meter or test strips may be working incorrectly.
You drop the meter.
You have repeated a test and the test results are still lower or higher than expected.
You are practicing the test procedure.
Note:
Professional users are instructed to follow federal, state, and local guidelines.